The HVD-100 is a type of wideband voltage divider which is designed to measure single-ended high voltage. The HVD-100 can measure multiple singleended signal such as DC, AC and pulse. The HVD-100 scaled down its input signal and outputs low voltage, which can easily be processed by universal instruments, such as oscilloscope, digital multimeters, etc. You can measure DC voltages up to 100 kV RMS and pulses up to 160 kV peak. The 20 MHz bandwidth enables you to measure fast, high-voltage signals.
YES
10 Days
12 months
| Parameter | Test conditions | Minimum | Typical values | Maximum |
| Standard Divider Ratio | 10000 : 1 | |||
| Max Peak Voltage (DC,AC) | 100kV | |||
| Max Peak Voltage (Pulse) | 160kV | |||
| Accuracy (DC - 2 Hz) | @ 25°C | 0.1% of range | ||
| Accuracy (2Hz - 1MHz) | @ 25°C | 1% of range | ||
| Accuracy (> 1MHz) | @ 25°C | 3% of range | ||
Bandwidth (-3dB) | DC-20MHz | |||
Nonlinear error | 0.022% | |||
| Rise time | 16ns | |||
| Input Impedance | 1.6GΩ / 25pF | |||
| Input Signal Type | single-ended | |||
| Output Signal Type | single-ended |
| Operation temperature | -15°C ~ 70°C |
| Storage temperature | -25°C ~ 80°C |
| Input connector | 4mm Banana plug |
| Output connector | BNC (jack) |
| BNC cable length | 9m |